Dionex ICS-1100 Ion Chromatography System
118
Doc. 065289-03 10/12
3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated
3.15 amps (P/N 954745). Thermo Fisher Scientific recommends always
replacing both fuses.
4. Reinsert the fuse holder into its compartment and push in until the tab clicks
in place.
5. Reconnect the main power cord and turn on the power.