5 • Service
Doc. 065289-03 10/12
117
8. Close the waste valve.
Do not overtighten.
9. Select the flow rate required for the analysis.
10. In the
Pump Settings
window, click
Pump Off
.
5.18 Changing Main Power Fuses
1. Turn off the main power switch.
2. The fuse holder is part of the main power receptacle (see
) on the
Dionex ICS-1100 rear panel. To remove the fuse holder, squeeze the tab on
the top of the holder to release it and pull the holder straight out of its
compartment.
Isopropyl alcohol may damage some columns. Be sure to thoroughly
rinse the alcohol from the pump, using the process described in
HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the Dionex ICS-1100.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau arrière du Dionex ICS-1100.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des Dionex ICS-1100.
Figure 5-26. Main Power Fuse Holder
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...