4 • Troubleshooting
Doc. 065289-03 10/12
73
•
Flow rate is set to 0
Set the flow rate from the Dionex ICS-1100 Control panel in Chromeleon.
•
Eluent valve is closed
Open the eluent valve from the Dionex ICS-1100 Control panel in
Chromeleon.
•
Pump not primed
•
Broken pump piston
Replace the piston (see
4.7
Erratic Flow/Pressure Reading
•
Pump needs priming
•
Damaged piston seal
Replace the piston seal (see
•
Dirty pump check valve
Clean or replace the check valve (see
•
Leaking liquid lines or fittings
Check the liquid lines and fittings for small leaks. Tighten or, if necessary,
replace the liquid line connection (see
). Refer to
Installation of
Dionex Liquid Line Fittings
(Document No. 031432) for tightening
requirements. The manual is included on the Thermo Scientific Reference
Library DVD (P/N 053891).
4.8
Excessive System Backpressure
•
Restriction in the system plumbing
Check all liquid lines for crimping or blockage. Make sure the ferrule fittings
are not overtightened onto tubing. Refer to
Installation of Dionex Liquid Line
Fittings
(Document No. 031432) for details. The manual is included on the
Thermo Scientific Reference Library DVD (P/N 053891).
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...