Dionex ICS-1100 Ion Chromatography System
56
Doc. 065289-03 10/12
•
Check the waste container daily and empty when needed.
Weekly
•
Once a week, check fluid lines for crimping or discoloration. Replace any
pinched lines. Replace damaged lines.
•
Check the junctions between the pump heads and the pump casting for
evidence of liquid leaks. If the piston seal wash tubing is not connected, check
the drain tubes at the rear of the pump heads for evidence of moisture. Normal
friction and wear may gradually result in small liquid leaks around the piston
seal. If unchecked, these leaks can gradually contaminate the piston housing,
causing the pump to operate poorly. If leaks occur, replace the piston seals
(see
•
Check the end-line filter (P/N 045987) and change if needed. When new, end-
line filters are pure white. If the system is in continuous operation, change the
end-line filter weekly, or whenever it becomes discolored. Replace the filter
more often if bacterial buildup is visible or if the eluent does not contain
solvent.
NOTE It is especially important to regularly replace end-line
filters when using aqueous eluents, which may
contaminate the filter with bacteria or algae. The
bacterial buildup may not be visible.
Every Six Months
•
Calibrate the cell (see
•
Calibrate the vacuum degas assembly (see
•
Replace the pump piston rinse seals and piston seals (see
Yearly
•
Thermo Fisher Scientific recommends performing preventive maintenance
annually, as well as before scheduled Performance Qualification tests. A
Dionex ICS-1100 Preventive Maintenance Kit (P/N 057954) is available for
this purpose.
•
Rebuild the auxiliary valve, if installed (see
•
If a Dionex AS Autosampler is installed, perform the preventive maintenance
procedure. The Dionex AS Preventive Maintenance Kit (P/N 060581) is
available for this purpose.
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...