5 • Service
Doc. 065289-03 10/12
87
5.1.4
Calibrating the Vacuum Degas Assembly
When to Calibrate
•
After every 6 months of use
1. Open the Dionex ICS-1100 Wellness Panel in Chromeleon (see
2. On the Wellness Panel under
Degas Calibration
, select
Calibrate
.
The degas pump runs for 90 seconds to allow a vacuum to be created.
After 90 seconds, the vacuum degas pressure reading is recorded.
When the vacuum degas assembly is running in monitor mode, this
calibration value is used to determine when to turn on the degas
pump.
3. Click
Log
to record the new calibration value in the Audit Trail.
5.2
Isolating a Restriction in the Liquid Lines
A restriction in the liquid plumbing will cause excessive system backpressure.
1. Begin pumping eluent through the system (including the columns).
2. Follow the flow schematics in
through the system, beginning at the suppressor
Regen Out
port. One at a
time, loosen each fitting and observe the pressure. The connection at which
the pressure drops abnormally indicates the point of restriction.
NOTE The numbers on the flow schematics indicate the order
in which liquid flows through the system components.
If the restriction has caused such high pressure that the system cannot be
operated, you must work forward through the flow schematic, adding parts
one at a time until an abnormal pressure increase (and hence, the restriction) is
found.
3. If the restriction is in the tubing or fitting, remove the restriction by back
flushing or by replacing the tubing or fitting (see
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...