D • FAQ
Doc. 065289-03 10/12
141
D.13 How do I clean a column?
See the troubleshooting section of the column manual.
D.14 Why is the conductivity high?
Possible reasons for high conductivity include:
•
The suppressor is not on. Turn on the suppressor from the Chromeleon
Control panel.
•
The suppressor needs regeneration. See the suppressor manual for
troubleshooting information. The suppressor manual is included on the
Thermo Scientific Reference Library DVD (P/N 053891).
•
The cell is out of calibration. See
for calibration instructions.
•
See
for additional troubleshooting information.
D.15 How do I configure and operate the auxiliary valve?
See
Installing the ICS-1100/ICS-1600/ICS-2100 Auxiliary Valve
(Document No.
065288), provided in the Auxiliary Valve Kit.