Primary Displays & Menus
The Total Menu
Note
The value for the low signal cutoff should usually be set at
approximately half of the value of the signal strength present under
flow conditions. Typically, signal strength is not significantly affected
by flow rate.
▲
From the Low Signal Action setup menu, select one of the following:
●
Zero: The flowmeter drops the reading to zero during LOS
condition.
●
Hold: The flowmeter holds the last valid reading during LOS
condition for about three seconds.
Figure
5–19.
Low Signal Action setup menu
The Total Menu
Access the Total menu to configure the totalizer parameters.
●
Totalizer Units setup menu (36)
The flow units elected for the totalizer display may be different from
the flow unit selected for the flow rate display.
Figure
5–20.
Totalizer Units setup menu
●
Totalizer Multiplier setup menu (37)
The totalizer value can be displayed with one of several multiplier
values. For example, 700 liters can be displayed as ‘700’ if the selected
multiplier value is X1, or it can be displayed as ‘7’ if the selected
multiplier value is X100.
Figure
5–21.
Totalizer Multiplier setup menu
Thermo Fisher Scientific
DCT6088 User Guide
5-7
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...