Operating & Configuring the Flowmeter
Quick Setup Configuration
Figure
4–6.
Pipe OD menu
If you know the pipe circumference but not the OD, enter the
circumference value instead and press
ENTER
and
the
down
arrow.
The screen shown in Figure 4–7 is displayed. Select
Circum
.
The flowmeter will calculate the pipe OD automatically.
Figure
4–7.
Pipe OD type options
b.
Select the Pipe ID setup menu (12), and enter the pipe ID. Press
Enter
.
Figure
4–8.
Pipe ID setup menu
c.
Select the Pipe Material setup menu (13). Press the
up
or
down
arrow
to scroll through the available options.
Note
Select
OTHER
if the material is not listed. You must then
enter the pipe sound speed (14) and pipe inside roughness (15).
▲
Figure
4–9.
Pipe Material setup menu
Note
Refer to
“The Pipe Menu”
(Chapter 5) for additional items not
addressed in the quick setup procedure.
▲
3.
If there is a liner, continue with this step. Otherwise, skip to step 4.
a.
Access the Liner Material setup menu (16). The screen shown in
Figure 4–10 is shown. Select one of the available options.
Thermo Fisher Scientific
DCT6088 User Guide
4-7
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...