Wiring
Power Connections
Figure
3–1.
Recommended cable routing
Power
Connections
ower
Connections
Warning
To prevent damage to the instrument, verify that the voltage to
be connected matches the voltage rating of the flowmeter. The voltage
rating is indicated beneath the power input terminals.
▲
Warning
Power connections must be made in accordance with local
standards or codes of practice.
▲
The power input terminals shown in Figure 3–1 can be connected to one
of the following voltages:
●
120 Vac, 50/60 Hz (setting should be on 110)
●
240 Vac, 50/60 Hz (setting should be on 220)
●
12 to 24 Vdc (setting should be on 110)
To connect power, remove the sheet metal access cover. Locate the
individual power input terminals. They are marked directly on the main
board or on a sticker attached to the main board. Connect the power
according to one of the following sections.
3-2
DCT6088 User Guide
Thermo Fisher Scientific