Chapter 6
Emergency Override & Master
Erase
Since the two procedures described below allow critical data to be accessed
and changed, this chapter may be removed from the manual to prevent
unauthorized use of these features.
Emergency
Overrides
rgency
Overrides
If a user-entered password is forgotten, the following emergency override
passwords maybe used: 42 for the system password and 43 for the scale
factor password. These override passwords may not be changed or disabled.
Performing a
Master Erase
Performing a
Master Erase
Caution
The master erase function erases all user-entered data and data in
the data logger. Note all configuration settings and download the data log
file if a record is desired prior to performing a master erase.
▲
To perform a master erase:
1.
Turn the flowmeter off and back on.
2.
When the message INITIALIZING... is displayed, press
ERASE
within
three seconds.
3.
The Master Erase screen is displayed. To continue, press the
5
key
within three seconds.
Figure
6–1.
Thermo Fisher Scientific
DCT6088 User Guide
6-1