System Installation
Transducer Mounting Methods
14.
Remove the paper from the pipe and trace the shape of the 9 o’clock
transducer in the same manner you did for the 3 o’clock transducer.
Ensure that the inside edge of the transducer (opposite the cable
connection) is even with the point just marked on the 9o’clock side of
the pipe.
Figure
2–18.
Z mounting method, step 14
15.
Mount the transducers with pipe straps by following steps 5-6 in
“Spacing & Mounting the Transducers”
earlier in this chapter.
Figure
2–19.
Z mounting method, step 15
The figure below illustrates the final Z method installation.
Figure
2–20.
WV and WW
Methods
For applications with pipe diameters smaller than 2 inches (50 mm), the
WV and WW methods are options that allow higher accuracy and stability
to be achieved when reasonable signal strength can be obtained (10% or
higher).
2-12
DCT6088 User Guide
Thermo Fisher Scientific