Hazardous Area Installation
European Certification
The transducers are certified IS by LCIE for EEx ia IIB T6 when IS
barriers are installed. The table below lists the European hazardous area
installation requirements.
European
Certification
Table
8–2.
European Hazardous Area Installation Requirements
Item
Zone 0
Zone 1
Zone 2
Unclassified
Instrument
N/A
EExd1 enclosure EExd1 enclosure IP65 enclosure
Transducers
EEx ia, barriers
required
EEx ia, barriers
required
EEx ia, barriers
required
Barriers not
required
Installation
drawing
2
Figure 8–3
Figure 8–3
Figure 8–3
N/A
1
Instrument should be installed in accordance with required codes, including use of explosion proof seals
for the wiring connections to the enclosure.
2
The flowmeter must be installed and wired in accordance with the specified installation drawing.
8-4
DCT6088 User Guide
Thermo Fisher Scientific