Primary Displays & Menus
The Relays Menus
12.
Enter a flow rate which equals the 4 mA (minimum anticipated)
reading, and press
Enter
.
Figure
5–50.
4 mA Span screen
13.
Press the
down arrow
, and enter a flow rate that equals the 20 mA
(maximum anticipated) reading. Press
Enter
.
Figure
5–51.
20 mA Span screen
14.
Access the Flow/Net Totalizer primary display (00) to complete the
current loop calibration process.
Note
The current loop output freezes for approximately 30 seconds after
calibrating or testing.
▲
The Relays
Menus
The RELAYS menu contains the setup menus for programming, viewing,
and testing the instrument’s relay parameters. The relays may be used for
functions such as alarm, pump control, etc. Up to four relays may be
independently programmed for a variety of on/off conditions.
Alternatively, the relay can send a pulse to a remote device (for example, a
totalizer, sampler, or chlorinator) whenever the internal totalizer advances
by one unit.
Programming
To program a relay, follow the steps below.
1.
Access the Program Relays menu (71).
Figure
5–52.
Program Relays menu
5-20
DCT6088 User Guide
Thermo Fisher Scientific
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...