Primary Displays & Menus
The Diagnostics Menu
The Diagnostics menu contains various view-only menus which display
important diagnostic parameters that are currently used or calculated by the
flowmeter. These parameters are helpful when troubleshooting the
flowmeter.
The Diagnostics
Menu
●
Signal Strength/Margin display (90)
This screen displays the signal strength in percentage and the margin.
The signal strength value displayed is the average of the signal strengths
for the upstream and downstream transducers. Margin is an indicator
of signal quality and is generally greater than 50%. Signal strength is
generally greater than 3% under good measurement conditions.
Figure
5–62.
Signal Strength/Margin display
●
Delta Time/Fluid Sound Speed display (91)
Displays the value for DeltaT and the fluid sound speed as measured by
the flowmeter. DeltaT is the difference between the upstream and
downstream travel times, expressed in nanoseconds (ns).
Figure
5–63.
Delta Time/Fluid Sound Speed display
●
Reynolds Number/Profile Factor display (92)
Displays the Reynolds number and flow profile factor currently being
used by the flowmeter. The flow profile factor is calculated by the
flowmeter and used to determine the effect of the flow profile on mean
measured fluid velocity.
Figure
5–64.
Reynolds Number/Profile Factor display
5-26
DCT6088 User Guide
Thermo Fisher Scientific
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...