Chapter 2
System Installation
The enclosure should be mounted to a sturdy vertical surface such as a wall.
The enclosure can be directly mounted with screws or indirectly mounted
with mounting ears. The flowmeter has 5/16” metal threaded inserts
located in the corners of the enclosure. The inserts are located in mounting
wells that are sealed from the interior of the enclosure when the door is
closed, preventing moisture from entering the instrument through the
mounting holes.
Enclosure
Mounting
Direct Mount
Method
The enclosure can be directly mounted to a wall by inserting four 1/4"
screws into the mounting wells from the front of the enclosure. The screws
act as “through bolts” for securing the unit to the wall.
Mounting Ears
Method
The enclosure can be mounted to a flat, vertical surface using the optional
mounting ears. The mounting ears can be oriented vertically or
horizontally as follows.
1.
Screw the four mounting ears to the metal threaded inserts on the back
of the enclosure using the 5/16” screws provided in the mounting ears
kit.
2.
Attach the ears to the wall with standard mounting screws.
Figure
2–1.
Mounting ears method
Thermo Fisher Scientific
DCT6088 User Guide
2-1
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...