System Installation
Transducer Mounting Methods
8.
Remove the paper from the pipe. Place the fold and intersection mark
together again, and fold the paper exactly in half.
Figure
2–12.
Z mounting method, step 8
9.
Mark along the new fold.
Figure
2–13.
Z mounting method, step 9
10.
Draw a horizontal line along the pipe from the centerline of the 3
o’clock transducer position. Use a level to ensure that the line is level
with the top of the pipe. The line should be at least 3 inches (76 mm)
longer than the transducer spacing calculated by the UltraScan software
or via menu 25.
For example, if the software calculates the spacing as 14 inches (356
mm), draw a line 17 inches (432 mm) long.
Figure
2–14.
Z mounting method, step 10
2-10
DCT6088 User Guide
Thermo Fisher Scientific