Maintenance & Troubleshooting
Replacing the Current Loop Module
Replacing the
Current Loop
Module
placing the
Current Loop
Module
Warning
Prevent possible electrical shock and / or damage to the meter.
Disconnect power to the meter prior to performing this procedure.
▲
1.
Disconnect power from the flowmeter.
2.
Open the flowmeter door and remove the sheet metal access cover.
3.
Disconnect wiring to the module.
4.
Remove the two screws and four insulating washers from the lower
corners of the module. Each screw has one washer above and one
washer beneath.
5.
To prevent damage to the connector, grasp the module by all four
edges and slowly lift it straight out from the main board.
6.
Align the pins on the new module with the connector on the main
board and press the module straight in.
Caution
To ensure proper operation and avoid damage to the module,
align the pins on the module properly with the connector before
inserting the module into the flowmeter. Ensure that the screw holes in
the module align exactly with the mounting posts on the main board. It
is possible to incorrectly align the module by one pin and force the
holes to line up, causing damage to the module.
▲
See the figure below.
7-2
DCT6088 User Guide
Thermo Fisher Scientific
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...