System Installation
Transducer Mounting Methods
Figure
2–9.
Z mounting method, steps 2–3
4.
Obtain a continuous sheet of paper longer than the circumference of
the pipe. Calculator paper tape or thermal printer paper works well for
this.
5.
Fold one end of the paper across the pipe’s width to produce a clean,
straight edge.
6.
Line the fold of the paper up with the horizontal centerline of the 3
o’clock transducer.
Figure
2–10.
Z mounting method, steps 4–6
7.
Wrap the paper firmly around the pipe, and mark the intersection
point where the fold comes in contact with the rest of the paper.
Figure
2–11.
Z mounting method, step 7
Thermo Fisher Scientific
DCT6088 User Guide
2-9