System Installation
Small Pipe Applications
2-14
DCT6088 User Guide
Thermo Fisher Scientific
Table
2–5.
Pipe Sizes
Spacing
3.0 to 3.5 inches (76.2 to 88.9 mm)
0.50 inches (12.7 mm)
2.5 to 3.0 inches (63.5 to 76.2 mm)
0.44 inches (11.2 mm)
2.0 to 2.5 inches (50.8 to 63.5 mm)
0.38 inches (9.7 mm)
1.5 to 2.0 inches (38.1 to 50.8 mm)
0.32 inches (8.13 mm)
1.5 inches (38.1 mm ) and smaller
0.25 inch (6.35 mm)
Figure
2–22.
3.
Remove the adhesive protection paper to expose the tape strips. Place a
strip on each side of the surface along the line. Press the strips down to
ensure good adhesion.
Figure
2–23.
4.
Apply coupling compound to the space between the tape strips. The
optimum height of the compound layer is approximately half the
height of the tape strips.
Figure
2–24.