Wiring
Power Connections
To connect the power for 120 Vac operation:
120 Vac Operation
1.
Connect the
hot wire to the L1(+)
terminal.
2.
Connect the
neutral wire to the L2/N(-)
terminal.
3.
Connect the
ground wire to the GND
terminal.
Single Phase 240
Vac Operation
To connect the power for single phase 240 Vac operation:
1.
Connect the
live wire to the L1(+)
terminal.
2.
Connect the
neutral wire to the L2/N(-)
terminal.
3.
Connect the
ground wire to the GND
terminal.
Double Phase 240
Vac Operation
To connect the power for double phase 240 Vac operation:
1.
Connect the
one hot wire to the L1(+)
terminal.
2.
Connect the
other hot wire to the L2/N(-)
terminal.
3.
Connect the
ground wire to the GND
terminal.
12 to 24 Vdc
Operation
Note
For DC operation, set the 4 –20 mA current loop function to loop
powered mode to prevent ground loops.
▲
To connect the power for 12 to 24 Vdc operation:
4.
Connect the
positive wire to the L1(+)
terminal.
5.
Connect the
negative wire to the L2/N(-)
terminal.
Thermo Fisher Scientific
DCT6088 User Guide
3-3
Summary of Contents for DCT6088
Page 2: ......
Page 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 78: ...This page intentionally left blank ...
Page 94: ...This page intentionally left blank ...
Page 118: ...This page intentionally left blank ...