System Installation
Transducer Mounting Methods
11.
Measure the spacing from the inside edge of the 3 o’clock transducer,
and mark this on the pipe.
Figure
2–15.
Z mounting method, step 11
12.
Wrap the paper firmly back on the pipe. Have the point where the ends
of the paper come together line up with the horizontal line on the 3
o’clock side of the pipe. Ensure that the inside corner of the straight
edge of the paper is aligned with the mark made for the transducer
spacing. Tape the paper down, or have someone hold the paper in
place.
Figure
2–16.
Z mounting method, step 12
13.
Go to the other side of the pipe (9 o’clock position), and mark the pipe
at the point where the marked fold and the inside edge of the paper
length intersect.
Figure
2–17.
Z mounting method, step 13
Thermo Fisher Scientific
DCT6088 User Guide
2-11