Operation
Settings
Thermo Scientific
49iQ Instruction Manual
3-41
●
Cell A Intensity:
Displays the current intensity in cell A.
●
Cell B Intensity:
Displays the current intensity in cell B.
●
Lamp Temperature:
Displays the current lamp temperature. User
can adjust low and high alarm limits.
●
Lamp Temp Sensor Short:
Displays OK/Fail for lamp temperature
sensor short.
●
Lamp Temp Sensor Open:
Displays OK/Fail for lamp temperature
sensor open.
●
Lamp Current:
Displays the current lamp Current. User can adjust
low and high alarm limits.
●
Lamp Connection:
Displays OK/Fail for lamp connection.
●
Lamp Short:
Displays OK/Fail for lamp short.
●
Lamp Heater Current:
Displays the current lamp heater current.
●
Bench Temp Sensor Short:
Displays OK/Fail for bench temperature
sensor short.
●
Bench Temp Sensor Open:
Displays OK/Fail for bench temperature
sensor open.
●
Board Communication Failure:
Displays OK/Fail for board
communication status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5 V Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
2.5 V Ref. Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
3.3 Volts:
Displays current voltage readings. Alarm limits are
not changeable.
●
5.0 V Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
Lamp Power:
Displays current lamp power readings. Alarm
limits are not changeable.
●
15 Volts:
Displays current voltage readings. Alarm limits are not
changeable.
●
-15 Volts:
Displays current voltage readings. Alarm limits are
not changeable.
Summary of Contents for 49iQ
Page 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Page 168: ......
Page 242: ......
Page 260: ......
Page 268: ......
Page 280: ......
Page 281: ......