The internal ozone generator provides easy determination of zero,
precision, and Level 1 span checks. The ozone level produced is a function
of light intensity at 185 nm and gas flow. The light intensity is varied by
changing the current into the lamp. The gas flow is held constant by a
pressure regulator followed by a capillary.
The zero air source option is a convenient system for generating pollutant
free zero gas for O
3
monitoring requirements. This option includes the
pump, filter, filter holder, and Silastic tubing.
A 5-10 micron pore size, two-inch diameter PTFE element is available for
the 49iQ. This filter should be installed just prior to the SAMPLE
bulkhead. When using a filter, all calibrations and span checks must be
performed through the filter.
Ozonator
Zero Air Source
PTFE Particulate
Filter