Calibration
Calibration Procedure
4-10
49iQ Instruction Manual
Thermo Scientific
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Span Coefficient button will
show what the span coefficient will bet set to if the Calibrate button is
pressed. Pressing the Calibrate button will save the new coefficient and
calibrate the instrument.
4.
Enter the O
3
calibration gas concentration using the pushbuttons, and
then press the
Calibrate
button to calibrate the O
3
reading to the O
3
calibration gas.
5.
Record the ozone concentration as determined by the calibration
photometer and the corresponding analyzer response.