Servicing
Optional DMC Ozonator
7-44
49iQ Instruction Manual
Thermo Scientific
Use the following to replace the lamp.
Equipment required:
Ball-end hex drive, 3/32
1.
Remove the ozonator DMC from the instrument. Refer to “
DMC
Ozonator Removal
” on page
7-41
.
2.
Using a 3/32 ball-end hex drive (angle alignment of screws), unfasten
two #4-40 cap screws.
3.
Pull lamp out.
Figure 7–51.
Replacing the Lamp
4.
Replace lamp and reassemble in reverse order.
Lamp Replacement