Operation
Calibration
Thermo Scientific
49iQ Instruction Manual
3-9
The Calibration menu allows the user to calibrate the system, setup
automatic calibrations, and view calibration data. See Chapter 4
“
Calibration
” for further instructions on how to run a calibration.
The following screens show the calibration screens in single range mode
and dual or auto range mode. The dual and auto range modes have two O
3
span factors (high and low). This allows each range to be calibrated
separately. This is necessary if the two ranges used are not close to one
another. For example, a low O
3
range of 50 ppb and a high O
3
range of
20,000 ppb. For more information about range modes, see “
Range Mode
Selection
” on page
3-72
.
Home Screen>Calibration (single range mode)
Home Screen>Calibration (dual or auto range mode)
The Calibration menu contains the following information:
●
Calibrate Background:
Sets the O
3
reading to zero.
●
Calibrate Span Coefficient:
Sets the span coefficient when in single
range mode.
Calibration