Operation
Calibration
3-18
49iQ Instruction Manual
Thermo Scientific
The Manual Calibration screen adjusts the zero background or span
coefficient based on a user entered value. See Chapter 4, “
Calibration
” for
instructions on how to run a Manual Calibration.
The following screens show the manual calibration screens in single range
mode and dual or auto range mode. In dual or auto range modes, “High
Range” or “Low Range” buttons are displayed to indicate the calibration of
the high or low coefficient.
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(dual or auto range mode)
The Manual Calibration screen contains the following information:
●
Adjust Background:
Allows the user to manually adjust the zero
background.
●
Adjust Span Coefficient:
Allows the user to manually adjust the span
coefficient when in single range mode.
Manual Calibration