Operation
Settings
3-82
49iQ Instruction Manual
Thermo Scientific
The Compensation screen provides compensation for any changes to the
instrument’s output signal due to internal instrument temperature, and
pressure variations.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
●
Temp Compensation:
Toggles temperature compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on
the analyzer's subsystems and output have been empirically determined.
This empirical data is used to compensate for any changes in
temperature.
●
Pressure Compensation:
Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to bench pressure variations. The effects
of bench pressure changes on the analyzer's subsystems and output have
been empirically determined. This empirical data is used to compensate
for any change in bench pressure.
●
Standard Temperature for Gas Units:
Allows the user to enter a value
between 0 and 30
°
C to correct the
g/m
3
, mg/m
3
, and g/m
3
concentration calculations.
Compensation
Summary of Contents for 49iQ
Page 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Page 168: ......
Page 242: ......
Page 260: ......
Page 268: ......
Page 280: ......
Page 281: ......