Operation
Settings
3-44
49iQ Instruction Manual
Thermo Scientific
●
Lamp Heater Current:
Displays the current lamp heater Current
reading.
●
Board Communication Failure:
Displays OK/Fail for
communication status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
2.5 V Ref. Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
3.3 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
5.0 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
5-20 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
15 V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
●
-15 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
24 V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
Note
If both the low alarm and high alarms are set to zero, then no alarm
will show.
▲