Servicing
49iQ Photometer DMC
7-34
49iQ Instruction Manual
Thermo Scientific
Use the following procedure to remove the pre-amplifier/detector assembly.
Equipment required:
Hex drive, 7/64
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug electrical cable from detector 1 and detector 2.
Figure 7–40.
Preamp Detector Cover Removal
3.
Using a 7/64 hex drive, unfasten two #6-32 cap screws.
Detector Assembly
Replacement