Troubleshooting
Troubleshooting Guide
Thermo Scientific
49iQ Instruction Manual
6-5
Problem
Possible Cause
Action
If using a sample particulate filter,
make sure it is not blocked.
Disconnect sample particulate filter
from the sample bulkhead; if flow
increases, replace filter.
Alarm – Cell A
Intensity
Alarm – Cell B
Intensity
Pre-amp gain not set
properly
Check gain adjustment.
Defective photometer DMC
PCB board
Replace photometer DMC PCB board.
Alarm – Zero Check
(Option)
Instrument out of
calibration
Recalibrate instrument.
Check gas supply.
Perform a manual calibration.
Alarm – O
3
Concentration
Concentration has
exceeded range limit
Check to ensure range corresponds
with expected value. If not, select
proper range.
Concentration low
Check user-defined low set point, set
to zero.
Alarm – Lamp Heater
Current (Photometer)
Defective heater
Replace heater as needed.
Alarm – Lamp Temp
Sensor Short
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Temp
Sensor Open
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Open
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Short
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp
Connection
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Short
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Heater
Current (Ozone)
Defective heater
Replace heater as needed.
Summary of Contents for 49iQ
Page 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Page 168: ......
Page 242: ......
Page 260: ......
Page 268: ......
Page 280: ......
Page 281: ......