Operation
Settings
3-60
49iQ Instruction Manual
Thermo Scientific
The Preventive Maintenance screen reminds the user when certain
instrument components need to be serviced/replaced. When the “Months
Left” has decreased to 2, the row is highlighted yellow. If the “Months
Left” is 1 or less, the row is highlighted red and the maintenance icon
(gears) will appear in the status bar located at the bottom of the screen.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Preventive Maintenance Alerts
The Preventive Maintenance screen contains the following information:
●
Across:
●
Component:
Device to be routinely serviced or replaced.
●
Interval in Months:
Expected period of time before a component
needs to be checked and/or serviced.
●
Months Left:
Count down timer in months. Remaining time since
the beginning of the service interval. When the value is 1 or less,
the row will be highlighted and it is suggested that the component
should be checked and/or serviced.
●
Reset:
Once the component is serviced/replaced, the user presses the
Reset button and the “Months Left” value resets to the “Interval in
Months” value.
●
Alert:
Allows the user to opt out of receiving preventive
maintenance alerts. Displays Enabled/Disabled for each
component.
●
Down:
●
Photometer DMC:
Service interval for photometer module
components.
Preventive Maintenance
Summary of Contents for 49iQ
Page 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Page 168: ......
Page 242: ......
Page 260: ......
Page 268: ......
Page 280: ......
Page 281: ......