Servicing
Solenoid Replacement
Thermo Scientific
49iQ Instruction Manual
7-39
Use the following procedure to replace the solenoids and optional solenoid.
(
Figure 7–45
).
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug electrical from Step POL board, solenoid valve sample (J8),
solenoid valve reference (J6), and optional solenoid valve sam/cal (J5).
3.
Disconnect 2X plumbing.
4.
Using a #2 Phillips drive, unfasten four captive hardware and pull out
the solenoid tray.
5.
Pull solenoid tray outwards.
Figure 7–45.
Removing the Solenoid Tray pt 1
Solenoid
Replacement