Calibration
Calibration Photometer System Preparation
4-6
49iQ Instruction Manual
Thermo Scientific
d
o
o
F
F
F
R
100%
A
R
A
A
E
1
2
1
where:
F
o
= Ozonator flow
F
d
= Dilutant zero air flow
E = Linearity error, in percent
A
1
= Assay of original concentration
A
2
= Assay of diluted concentration
R = Dilution ratio
Note
The inherent linearity accuracy of the 49iQ Primary Standard (or a
49iQ modified as described earlier) is greater than the accuracy
measurements of the mass flow meters.
▲
Use the following procedure to check the calculations are complete and
accurate:
1.
With the ozone generator in the manual mode, i.e., gain set to zero,
adjust ozone level to generate a level in excess of 0.5 ppm. Wait until
ozone concentration is stable.
2.
From the Home Screen, press Settings>Health Check>Status and
Alarms>
Intensity Check
. From the Intensity Check menu, select
Intensity A Reference Gas.
Wait for stable frequency reading and note
as Io(A). Press the
button to return to the Intensity Check screen.
3.
From the Intensity Check screen, select Intensity A Sample Gas. Wait
for stable frequency reading, note as I(A). Press the
button to return
to the Intensity Check screen.
4.
From the Intensity Check screen, select Intensity B Reference Gas.
Wait for stable frequency reading, note as Io(B).
Summary of Contents for 49iQ
Page 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Page 168: ......
Page 242: ......
Page 260: ......
Page 268: ......
Page 280: ......
Page 281: ......