calibrating sensors
chapter 5
5-7
5-6
Note:
A full 2-point tonometered gas calibration procedure is highly recommended to
get the best measurement performance. Use of default values should be limited to
unusual circumstances which preclude the ability to do the normal calibration proce-
dure. In any case, it is highly recommended that you enter the K
+
calibration value
from the sensor pouch to ensure optimal K
+
measurement performance.
If the last calibration values are suspect for any reason, it is recommended that
you change the default setting to “Factory default values” before entering Operate
mode.
Note:
You must turn the monitor off and back on for a change in calibration status to
take effect.
Note:
If you intend to turn off the monitor after calibration and then power it back up
before operation, you should use the “Last calibration values” setting. This will ensure
that the calibration values will be retained upon power up.
Note:
An aborted calibration will not affect stored “Last calibration values” as defaults.
When the calibration is finished, the calibration results are printed (if the printer is
active), the values are stored, and the following screen appears (if the calibration was
successful).
Note:
If there are any error messages, consult Chapter 9 “Troubleshooting,” to help
resolve the problem.
13:26
9-25-18
You can press this
key to perform a
calibration
verification, if en-
abled in setup.
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...