troubleshooting
chapter 9
9-23
CF error code is for Calibrator (CDI 540)
Error code
(Recoverable)
Description
Cause
Recommendation
CF01
Detected at
calibration start up.
Calibrator I2C failure.
Exit calibration and contact Terumo
CVS Technical Service.
CF02
Detected at
calibration start up.
Calibrator ADC failure.
Exit calibration and contact Terumo
CVS Technical Service.
CF03
Detected at
calibration start up
Calibrator reference voltage
error.
Exit calibration and contact Terumo
CVS Technical Service.
CF04
Detected at
calibration start up.
Bottle A valve failure or
Calibrator leakage or bad
transducer for bottle A.
Exit calibration and contact Terumo
CVS Technical Service.
CF05
Detected at
calibration start up.
Bottle B valve failure or
Calibrator leakage or bad
transducer for bottle B.
Exit calibration and contact Terumo
CVS Technical Service.
CF06
Detected at
calibration start up.
Bad exhaust valve or mix-
line transducer
Exit calibration and contact Terumo
CVS Technical Service.
CF07
Detected at
calibration start up.
Mix line transducer failure
or bottle A, B valve failure.
Exit calibration and contact Terumo
CVS Technical Service.
CF08
Detected at
calibration start up.
Bad bottle A pressure
regulator or bad mix line
transducer.
Exit calibration and contact Terumo
CVS Technical Service.
CF09
Detected at
calibration start up.
Bottle A valve failure, bottle
A pressure regulator failure,
large leak.
Exit calibration and contact Terumo
CVS Technical Service.
CF0A
Detected at any
time during
calibration.
Bad exhaust valve.
Exit calibration and contact Terumo
CVS Technical Service.
CF0B
Detected at
calibration start up.
Bad bottle B pressure
regulator or bad mix line
transducer.
Exit calibration and contact Terumo
CVS Technical Service.
CF0C
Detected at any
time during
calibration.
Bad calibrator memory or
un-calibrated calibrator.
Restart calibration process and see
if the error code persists. If the
error code persists, exit calibration
and contact Terumo CVS Technical
Service.
CF11
Detected at
calibration start up.
Bottle B valve failure, bottle
B pressure regulator failure,
large leak.
Exit calibration and contact Terumo
CVS Technical Service.
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...