troubleshooting
chapter 9
9-11
Message/Problem
Meaning and Corrective Action
Arterial (or Venous
or H/S) module
failure: ________
Measurements
from module will
be discontinued.
Contact Terumo
Technical Service.
This message refers to a detected module hardware fail-
ure or communications failure with the monitor. This may
occur due to sudden component failure or physical dam-
age to the module, monitor or cable. An identifying error
code appears in the blank. Record this error code for later
reporting to Terumo Technical Service. You may continue
using the system although measurements from this mod-
ule will be discontinued and will no longer be displayed.
Contact your Terumo Technical Service representative to
arrange for repair.
Monitor failure
detected:
________
Contact Terumo
Technical Service.
This message refers to a detected monitor hardware fail-
ure. This may occur due to sudden physical damage to
monitor, component failure, or failure of all modules. An
identifying error code appears in the blank. Record this
error code for later reporting to Terumo Technical Ser-
vice. You will not be able to continue use of the system.
Contact your Terumo Technical Service representative to
arrange for repair.
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...