quick start
chapter 2
2-3
2-2
Installation
See Chapter 6 for installation warnings and more details on installation.
1
If the Terumo CDI H/S Cuvette is in a single
sterile package, remove it and insert it into the
extracorporeal circuit using aseptic technique.
2
Prime the circuit.
3
Install the CDI Shunt Sensor into the circuit us-
ing aseptic technique:
Place the cable-head on the cable-head
bracket.
Remove the top white luer cap and attach
the top of the sensor to the purge/shunt line
in the circuit.
Remove the sensor filter/sparger assembly
using aseptic technique and attach the end
of the purge/shunt line to the bottom of the
sensor.
4
If you are using a shunt bypass line, install the
CDI Shunt Sensor into the shunt bypass line
using aseptic technique:
Stop the pump and/or clamp off the 1/4 inch legs of the shunt bypass line.
Remove the blue cap from the female end of the shunt bypass line and the
small top luer cap (white) from the top of the CDI Shunt Sensor. Attach the
CDI Shunt Sensor to the female end of the shunt bypass line.
Remove the white cap from the male rotating luer connector on the shunt
bypass line. Remove the filter/sparger assembly (clear) from the CDI Shunt
Sensor. Attach the male rotating luer end of the bypass line to the CDI Shunt
Sensor, and securely tighten the rotating luer onto the CDI Shunt Sensor.
Unclamp the shunt bypass lines.
5
Prime and debubble the CDI Shunt Sensor(s), inspecting the purge/shunt line,
the shunt bypass line, and the CDI Shunt Sensor for bubbles.
6
Attach the CDI H/S Probe to the CDI H/S Cuvette:
Remove the CDI H/S Probe from
the probe holder.
Connect the bottom of the CDI
H/S Probe to the bottom of the
H/S Cuvette.
Press the CDI H/S Probe and
CDI H/S Cuvette together until
the release lever latches.
5000605
5000206
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...