communication with other devices
appendix c
C-15
CDI Pump Interface.
The CDI Pump Interface is a 2-way communications protocol
designed for future pumping systems, including the Terumo® Advanced Perfusion
System 1. It provides a flexible format to allow the System 550 to get flow data as
desired from the pump and for the pump to request data as desired from the monitor.
The interface protocol is fully described later in the appendix section.
Pump interfacing tips
Any pump interface cable used is specific to each pump and should be ordered from
Terumo Customer Administration. The interface cable will always connect to the
pump interface port (only) on the monitor. When an interface cable is not connected
to the pump interface port, use the protective cap to cover the port. The data output
port (located just above the pump interface port) is reserved for sending data to com-
puters or data acquisition systems.
Note:
Please consult the Operator’s manual of the specific pumping system being
used for instructions on how to properly setup the pump for sending data through its
serial interface.
Pumping System
Special Instructions
Terumo® Advanced
Perfusion System 1
Interfacing with the Terumo® Advanced Perfusion System
1 can only be done with the CDI interface module (Terumo
part number 803479). The System 1 Perfusion screen must
be configured to include the CDI module. See the System 1
Operator’s Manual for instructions on adding the CDI mod-
ule to a perfusion screen configuration. The physical con-
nection between the CDI module of the System 1 to the CDI
System 550 is the same process as the CDI System 500.
Make sure the correct interface cable is used (Terumo part
number 804981), and that it is connected in the proper
orientation (see labels on cable to determine correct orien-
tation).
Set up the CDI System 550 as follows: use the system
mode select key and the direction keys to access screen
number 5 in setup mode, the calculations screen. Set the
source to pump and the pump type to Terumo CDI Pump
Interface.
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...