setting up the hardware
chapter 3
3-1
PB
This chapter explains how to set up the Terumo Cardiovascular Systems CDI
Blood Parameter Monitoring System 550 Monitor, printer, and Terumo CDI Model
540 Calibrator, and describes the process of powering up the monitor.
Setting up the monitor
To set up the monitor, follow these steps:
1
Attach the monitor pole clamp securely to a heart-lung machine pole.
Tighten the pole clamp until it’s secure.
2
Attach the CDI System 550 Monitor to the tray of the monitor pole clamp.
Place the monitor on the tray, positioning it correctly on the alignment cone,
then pushing up while turning the locking screw on the bottom of the tray until
it’s tight. Loosen the adjustment knob and adjust the monitor to the desired
position. Tighten the adjustment knob to secure.
When you want to remove the monitor from the tray, you can loosen the locking screw
until the monitor is no longer locked to the tray.
Warning
•
Never leave the monitor on the pole clamp tray without securing the locking
screw.
•
Do not roll equipment over the cables of the CDI System 550. Mishandling
can cause damage and deterioration of system performance.
5000303
Alignment cone
Locking screw
Adjustment knob
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...