system specifications
appendix b
B-4
System accuracy limits
The CDI System 550 has been subjected to rigorous bench tests to simulate the
clinical use of the device and assessed its accuracy and precision over the system
operating ranges of the measured parameters. Blood samples taken from the test
circuit were analyzed in conventional analyzers, and these results were compared (on
a sample-by-sample basis) to analyses provided by the CDI System 550. Accuracy
values were generated by using the results from the CDI System 550 minus conven-
tional analyzer results The following table shows the mean difference between the
two measurement techniques, and the standard deviation of the distribution differ-
ences found.
pH Sensor (pH units):
Total Hemoglobin Value (g/dl):
Mean: 0.007
Mean: -0.17
Standard Deviation: 0.014
Standard Deviation 0.54
PCO
2
Sensor (mmHg):
Potassium Sensor (mmol/L):
Mean: 0.0
Mean: 0.05
Standard Deviation: 2.9
Standard Deviation: 0.18
PO
2
Sensor − Arterial (>80 mmHg):
Hematocrit Value (%):
Mean: -0.5
Mean: -0.5
Standard Deviation: 5.5
Standard Deviation: 1.7
PO
2
Sensor - Venous (<80 mmHg):
Mean: -0.4
Standard Deviation: 0.8
Oxygen Saturation Value (%):
Mean: -0.5
Standard Deviation: 1.6
Summary of Contents for CDI 550
Page 29: ...chapter 2 2 6 This page is intentionally left blank...
Page 39: ...chapter 3 3 10 This page is intentionally left blank...
Page 53: ...chapter 4 4 14 This page is intentionally left blank...
Page 85: ...chapter 8 8 4 This page is intentionally left blank...
Page 109: ...chapter 9 9 24 This page is intentionally left blank...
Page 113: ...chapter 10 10 4 This page is intentionally left blank...
Page 164: ...patent information F 1 Patent Information U S Patents 5 296 381 5 508 509 5 591 400...
Page 165: ...appendix f F 2 This page is intentionally left blank...
Page 169: ...appendix g G 4 This page is intentionally left blank...