Section G – Dissolution Testing
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
88 of 130
EAR99 technology subject to restrictions on cover page.
Events tab
The Events tab allows the user to add events or modify the
Event Table.
The Event Table is a sequence of events that will take place at
user-programmed times. There are three types of events:
sampling, set parameter and action. Sampling events must
have a time assigned to them. Events may occur during the
test, pre-run (before the test), or post-run (after the test).
NOTE: Pre-run and post-run events cannot be sampling
events.
Pre-Run: Events programmed as a pre-run event will occur
before the test begins.
For example, an operator programs a pre-run event of Drive
Speed for 25 rpm. Prior to the test starting the motor will spin
at 25 rpm. When the test begins, the motor will go to the speed
programmed on the Dissolution tab.
Post-Run: Events programmed as post-run will occur after the
test has completed.
For example, an operator programs a post-run event of Bath
Temperature for 40 °C. Once the test completes, the bath
temperature will be set to 40 °C.
To edit or delete an event, touch the field in which it appears;
this will open the Edit Event screen.
To delete all the events in a method, touch the Delete All
button.