Section E – Autosampler Installation
Location
This section provides information on determining if a location
is suited to a CD AutoPlus and AutoFill.
Environmental Requirements
The CD AutoPlus samplers should be placed one on the left
(Side A) and one on the right (Side B) of the CD14 Tester. The
Teledyne Hanson AutoFill collectors should be placed on top
of each CD AutoPlus.
The location should be clean and free of any issues which may
influence dissolution testing. Issues that may affect dissolution
results include vibration, significant heat, or significant cold.
Space Requirements
The bench should be at least 60 cm (24 in.) deep and
clearance above the units of at least 82cm (32 in.).
The bench should be level and flat.
The bench should be capable of supporting weight given in the
specifications section.
For additional pieces of equipment (e.g., UV/Vis
spectrophotometers, etc.), see the user guides for the
respective pieces of equipment.
Electrical Requirements
Each CD AutoPlus requires a single grounded outlet within 1
meter (3 ft.) of the location of the system.
Each AutoFill requires a single grounded outlet within 1 meter
(3 ft.) of the location of the system.
The CD AutoPlus and AutoFill are available in one
configuration, 100 V through 240 V. The voltage is
automatically set when the voltage is supplied to the external
power supply.
Unpacking and Inspection
This section provides instructions on unpacking and setting up
each CD AutoPlus and AutoFill.
1.
Unpack each CD AutoPlus individually from its
shipping container.