Section H – CD14 Tester Maintenance
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
100 of 130
EAR99 technology subject to restrictions on cover page.
3.
Touch the Adj Temps button at the bottom of the
screen.
4.
Touch the probe you want to calibrate.
5.
Use the numeric keypad to enter the temperature
displayed on the NIST traceable thermometer, then
touch OK.
Moving and Storage – CD14 Tester
If the CD14 Tester will not be used for a short period of time (3
days or more), the power should be turned off. If the tester will
not be used for an extended period of time (30 days or more),
the instrument should be cleaned, the bath drained, and the
tester unplugged from the power source.
The instrument is heavy and great care should be taken when
moving it. If the instrument is moved to a new location,
revalidation should be considered.
Five people are required to lift the CD14 dissolution tester.
Lifting should be done with two people situated on each side of
the instrument. Lift the equipment by the base frame using the
screw-in handles provided.
Warning:
Lifting without assistance may cause injury.