Section G – Dissolution Testing
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
85 of 130
EAR99 technology subject to restrictions on cover page.
Vessel Goal Temperature: The desired temperature of the
vessels. This option is only applicable when the Digital
Temperature Probes field is set to Enabled. The range is 0
(off), 25.00 to 51.00 °C.
Sampling tab
Sampling Mode: determines the type of sampling procedure:
Manual: For manual sampling. When manual sampling
is selected, the user has the option of defining how
long the paddles remain spinning after a test has
ended, to allow for collection of the last sample. This
time is entered in the Vessel Offset Time field.
Automated (At Point): In At Point mode, the syringe
begins loading the sample volume (collect volume)
within a few seconds of the programmed sampling time
point. All rinse and backflush-filter operations (if any)
will occur before the programmed sample time.