Section G – Dissolution Testing
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
87 of 130
EAR99 technology subject to restrictions on cover page.
Magazine Travel: This parameter defines magazine stroke
from home to the sampling position. The range is 0 (disabled)
to 105 mm.
NOTE: The travel values may require fine tuning as necessary
to account for additional distance that may be needed due to
tip filters.
Media Replace: Volume from the media replace source
(typically Source C if configured) to be transferred to the
current source after sampling. The range is 0 – 99.9 mL.
Generally, the media replace volume should be set to be the
same as the total of the rinse volume and the collect volume.
Options tab
The Options tab allows for testing options to be set outside of
sampling and dissolution tester settings.
Number of Vessels: the number of vessels that will be used
during the test. The options are 6 or 12.
Perform Infinity Test: If checked, performs an infinity test with
the designated infinity test length and infinity test speed.
Infinity Test Length: The length of the infinity test which follows
the dissolution test. The range is 0:00 to 9:59 in h:mm format.
Infinity Test Speed: Sets the speed of the infinity test that
follows the normal test. The range is 0, 25 to 250 rpm.