CD14 Tester General Specifications
Spindle Speed:
Range: 25 to 250 rpm
Accuracy: ± 1 rpm
Displayed resolution: 0.1 rpm
Temperature:
Programmable temperature range: 25.0 to 51.0 °C
Accuracy: ± 0.5 °C
Control: ± 0.1 °C
Digital Temperature Probes (all probes):
Range: 10 to 60 °C
Accuracy: ± 0.1 °C from 30 – 55 °C
Displayed resolution: 0.01 °C
Minimum submersion depth: 25 mm
Response time: < 45 seconds (normally 30 sec.)
CD AutoPlus Specifications
Weight:
Dry: 21.3 kg (47.0 lbs.)
Size:
Height: 46.8 cm (18.4 in.)
Width: 38.1 cm (15.0 in.)
Depth: 66.0 cm (26.0 in.)
Power:
Input: 24 VDC, 3.75 A, 90 W
Only use external power supply (listed below)