Section D – Tester Installation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
37 of 130
EAR99 technology subject to restrictions on cover page.
Start the Pump
1.
Before starting the pump, the water level needs to
be at least to the fill line indicated on the bath.
2.
At the display, touch the center of the black ribbon.
This will bring up a keypad window.
3.
At the keypad window, touch the “Pump ON” button.
Any additional air that blows out into the bath will
stop shortly. Note: If water doesn’t start flowing
when the pump is started, then shut the pump off at
the displayed keypad, inject water into the two
bottom (left and right) ports within the bath a couple
of times with the supplied syringe. Start the pump
again.
Touch Here