Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
7.7 Use of Zero/Span Valves or IZS with Remote Contact
Closure
The Zero/Span valve or IZS options can be operated using Remote Contact Closures provided on
the rear panel. See Figure 2-2 for connector location and pinout. When the contacts are closed,
the analyzer will switch to zero or span mode. The contacts must remain closed for at least 1
second, and will remain in zero or span mode as long as the contacts are closed. If Dynamic
Zero/Span is enabled, the calibration is adjusted at the end of the zero or span time, otherwise
zero or span is just checked, not adjusted.
In order to do another remote check, both contact closures should be held open for at least 1
second, then may be set again. Table 7-9 shows what type of check is performed based on the
settings of the two contact closures.
Table 7-9: IZS or Z/S Valves Modes with Remote Contact Closure
Ext Zero CC
Ext Span CC
Operation
Contact Open
Contact Open
State when in SAMPLE mode, normal monitoring
Contact Open
Contact Closed
Span check or calibrate
*
Contact Closed
Contact Open
Zero check or calibrate
*
* Calibrate only if Dynamic Calibration is enabled.
7-11
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...