9.2.3 No Response to Sample
1.
Confirm correct operation of analog output by performing Analog Output Step Test in
2.
Confirm general operation of analyzer.
A.
Check for AC Power, Section 9.2.1.
B.
Do flow checks, Section 9.2.2.
C.
Confirm that sample gas contains NO or NO
2
.
3.
Check instrument electronics.
A.
Do ELEC TEST procedure in DIAGNOSTIC menu Section 9.1.3.2.
B.
Do OPTIC TEST procedure in the DIAGNOSTIC menu Section 9.1.3.3.
If the M200A passes ET and OT, that means the instrument is capable of detecting light
and processing the signal to produce a reading. Therefore, the problem is in the
pneumatics or ozone generator.
4.
Check ozone generator subsystem. Do diagnostic test of ozone generator subsystem, see
5.
Check for disconnected electrical cables to sensor module.
9-28
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...